Key Features
Auto Ranging
- Automatic control
- Internal solenoid for switching ranges
Data Recording
- USB enabled data recording
- Customizable reports
Portable & Compact
- Lightweight design
This unit has two flow ranges and two pressure channels. The SLM+ quantifies the component’s leakage rate by maintaining a constant gas pressure against the barrier being tested while measuring the gas flow rate needed in order to maintain that pressure constant. SLM+ is designed for instances where lower flow rates are observed. Using two ranges, a 25 to 1 turn-down ratio on each range is achieved with a 1% of full scale accuracy over each of the two ranges. Full scale flow ranges up to 20 slpm may be ordered.
24 VDC
Max. Supply Pressure: 150 psig
Max. Test Pressure: 100 psig
Accuracy: ±0.5% of Full Scale
Low Flow Range: 8 - 200 sccm
High Flow Range: 0.2 - 5 slpm
25 to 1 turn-down ratio on each range
Accuracy: ±1% of Full Scale